Toyota RAV4 manuals

Toyota Sienna Service Manual: Calibration

1. SELECT COMPASS DISPLAY MODE

  1.  The MODE switch allows you to select the Display or Non-display mode of the compass.

HINT: In compass display mode, the display indicates outside temperature/average fuel consumption/ instantaneous fuel consumption/distance to empty with compass.

2. SET ZONE

  1.  Deviation between the "magnetic north" and "actual north" differs depending on the location. Therefore, adjustment of the magnetism is required. Since the magnetic condition differs depending on the area where the vehicle is used, it is necessary for each user to set the zone (Refer to Zone Setting Map).

    The zone setting can be changed using the MODE switch of the accessory meter assembly.

3. PERFORM CALIBRATION

  1.  Because each vehicle has its own magnetic field, calibration should be performed for each vehicle.

    This compass function is used when storing the record of the vehicle's magnetic field.

4. WHEN COMPASS IS MAGNETIZED

  1.  A compass could be magnetized during shipping by vessels or freight cars. Therefore, make sure to perform calibration and ensure that calibration is performed properly before delivery. If it cannot be done (cannot be completed in spite of driving around several times), it may be caused by magnetization. Demagnetize the vehicle using a demagnetizer and perform calibration again.

5. SET COMPASS

SET COMPASS

6. ZONE SETTING MODE (PROCEDURE "A")

  1.  Pressing the E/M and MODE switches simultaneously for 5 seconds in the normal mode will activate the zone setting mode. A number (1 to 15) is displayed on the compass display.

    NOTICE: In the initial state, "VAR" is displayed.

  2.  The displayed number increases by 1 every time the E/M switch is pressed. Referring to the map, check the number for the area where the vehicle will be used and set the zone number.
  3.  Leave it untouched for several seconds after setting and check that the compass display shows an azimuthal direction (N, NE, E, SE, S, SW, W or NW) or "CAL".

7. CALIBRATION SETTING MODE (PROCEDURE "B")

  1.  Pressing the E/M and MODE switches simultaneously for 10 seconds in the normal mode will activate the calibration setting mode.
  2. Drive the vehicle at a slow speed of 8 km/h (5 mph) or less in the circular direction.
  3.  Driving around the circle 1 to 3 times will display the azimuthal direction on the display, completing the calibration.

NOTICE: After the calibration is completed, it is not necessary to perform the above procedures unless the magnetic field strength is drastically changed. If this happens, the azimuthal display will be changed to "CAL".

CALIBRATION SETTING MODE

Accessory meter
COMPONENTS ...

Removal
1. REMOVE ROOF CONSOLE BOX ASSEMBLY 2. REMOVE ACCESSORY METER ASSEMBLY  Using torx socket wrench (T10), remove the 5 screws and accessory meter assembly. ...

Other materials:

Installation
1. INSTALL HEATED OXYGEN SENSOR (for Bank 2 Sensor 2) (a) Using SST, install the heated oxygen sensor to the front exhaust pipe. SST 09224-00010 Torque: 40 N*m (408 kgf*cm, 30 ft.*lbf) for use with SST 44 N*m (449 kgf*cm, 32 ft.*lbf) for use without SST HINT: Use a torque wrench wit ...

Taking out the spare tire
Slide the front passenger’s seat to the frontmost lock position and the right side second seat to the rearmost lock position so they will not interfere with the jack handle. Open the right side sliding door and you will find the flap on the floor. Pull back the flap to find the ...

Evaporative Emission Control System Leak Detected
DTC SUMMARY DESCRIPTION The circuit description can be found in the EVAP (Evaporative Emission) System (See page ES-409). INSPECTION PROCEDURE Refer to the EVAP System (See page ES-412). MONITOR DESCRIPTION 5 hours*1 after the ignition switch is turned off, the electric vacuum pump c ...

© 2011-2026 Copyright www.tsienna.net